DEHAS QualityMIX – Precision medical gas mixers

Model Flow rate (l/min) Applications
High Flow 15-120 Intensive care, emergency care, air transport
Low Flow 3-30 Anaesthesia, low flow applications

Technical specifications: Available in High Flow (15-120 l/min) and Low Flow (3-30 l/min). Continuous precise mixing of medical gases (oxygen/air or oxygen/nitrous oxide). FIO2 accuracy ±3%. Integrated pneumatic check valve system, rotary FIO2 selector (21-100%), 100% O2 purge function, multiple outlets (with/without bleed valve).

Target patients: Infants, children, adults with hypoxaemia/respiratory failure in healthcare settings (intensive care, emergency, medical transport, anaesthesia, home care).

Clinical applications: Controlled oxygen therapy, intensive care, paediatrics, general anaesthesia, nitrous oxide analgesic sedation, emergency/aeromedical medicine, neonatal resuscitation, integrated flow meters.

Specific features: Robust stainless steel construction, MRI compatibility (3 Tesla), secure DISS/NIST connections. Pneumatic non-return system, rotary FIO2 selector, audible alarm (≥80 dB) for power supply failure. Multiple accessory options.

Patient benefits: Safe administration of precise oxygen concentration, minimising toxicity/hypoxaemia risks.

Clinician benefits: Reliable, repeatable control of gas flow/concentration, integrated alarms to prevent supply errors, ease of use, reduced maintenance (simple disinfection, no sterilisation required).